|

Overview of the Imaging Procedure
The image of NGC253 was captured using an SBIG ST-10XME and Celestron 14" SCT
operating at f/7 (with a Reducer/Corrector in place). The Reducer was used in order to provide a wide
enough field for the object and to allow shorter exposures. The image is
an LRGB composite. This requires shooting a set of images through the
clear filter to create a black-and-white luminance image, then shooting a set of
images through red, green, and blue filters to create a color image. The
color image is then combined with the luminance image to create a final image.
The reason for shooting an LRGB instead of a plain RGB is that it actually
allows for a shorter overall exposure time. This is because the RGB images are shot
with the CCD camera binned. This increases the sensitivity but lowers the
resolution. By then shooting the luminance image at full resolution then
combining it with the low-resolution color image produces a high-resolution
color image that actually requires less total exposure time. This is
discussed in more detail below. A total of 15 images were captured, plus 6
dark frames. The camera was self-guided by the built-in autoguider of the
ST-10XME.
Basic Steps
-
Polar align telescope
-
Attach CCD camera
-
Set CCD camera temperature
-
Balance telescope
-
Align computerized mount
-
Focus CCD camera
-
Find target
-
Focus telescope
-
Find guide star
-
Calibrate autoguider
-
Begin guiding
-
Start Luminance exposure sequence
-
Stop autoguider
-
Select Red filter
-
Focus telescope
-
Begin guiding
-
Start Red exposure sequence
-
Repeat previous 5 steps for Green and Blue
-
Take dark frames
Initial Setup
Basic setup includes balancing the telescope,
drift polar alignment, and
aligning the computerized mount, each detailed
in their respective sections on this site. This may be unnecessary for a
permanently-mounted telescope, but this setup was a portable instrument
transported to a dark site for imaging. For this setup, the telescope was
initially roughly polar aligned and balanced for visual use. Drift polar alignment
was done visually with a crosshair eyepiece. The mount was then powered
off, the eyepiece removed and the focal reducer and CCD camera installed, the
telescope re-balanced, the mount powered up again, the CCD turned on and started
cooling, and the mount computer aligned using the CCD. For aligning the
mount the CCD was only roughly focused as accurate focus was achieved at a later
step.
The CCD was cooled to -5°C. There are nice
things about living in Arizona, but one drawback is that cooling CCDs is
difficult. The ambient temperature while imaging NGC253 was around 70°F.
Imaging was done from about 11:00PM to 1:00AM in mid-September.

Above: C14, AP900, and ST-10 used for imaging NGC253
Focusing
The telescope was roughly focused for the purpose of aligning the
computerized mount. The scope was then pointed to the target, NGC253, and
a more precise focus was achieved. Focusing was done manually using a
Feathertouch SCT MicroFocuser and the Focus mode of MaxIm DL.
A 1-second exposure was used for focusing. Initial focusing was done
with the camera Binning set to 3. Then a subframe was selected and the
Binning set to 1. Focus was then achieved using the Inspect mode of MaxIm
DL by minimizing the FWHM value. Manual focusing is not as fast as using
an autofocuser, but in this case it took less than 5 minutes.

Above: Initial focus settings. For fine focusing Binning was set
to 1.

Above: Minimizing the FWHM value in the Inspect window
Finding a Guide Star
Finding a guide star is sometimes a bit tricky, although if you ever tried
finding a guide star for manual guiding using an off-axis guider during the days
of film astrophotography, you know it is infinitely easier than it used to be!
Software is very helpful here. For this image, Desktop Universe (now
incorporated into Starry Night Pro Plus) was used to determine the best position
for the camera in order to frame the galaxy properly and to locate a guide star.
NGC253 is large enough that it only fits into the field of view of this
telescope/CCD combination diagonally. On a German equatorial mount,
such as was used here, the default position of the guide chip in an SBIG
self-guiding camera (with the cables hanging down from the camera) is to the
east of the main CCD chip. This is shown below.

Above: Initial orientation of the CCD and guide chip
Any star between the two dotted circles is accessible to the guide chip
simply by rotating the camera. There are three relatively bright choices.
The bright star at the top is the best choice, in terms of brightness, but the
target galaxy would not fit in the field with the camera in that orientation.
The same is true of the potential guide star directly below the galaxy.
However, by convenient coincidence, the guide star to the lower left of the
galaxy (at about the 8 o'clock position) is perfect. When the camera is
rotated to place that star on the guide chip, the galaxy is framed perfectly.

Above: Selecting a guide star and framing the galaxy
According to Desktop Universe this is 160° from
the initial position. So the camera is rotated slightly less than halfway
around. The cables are now pointed up away from the dovetail plate
attaching the scope to the mount and slightly to the south, or right side if
looking at the back of the scope. If this seems complicated, that's fine:
I agree. So I made a cheat sheet for determining CCD orientation with a
German equatorial mount. I carry this with me into the field for imaging.
After finding the guide star, be sure to go back
and double check that your main target is framed properly.
Autoguiding
Once the guide star is located, the autoguider
needs to be calibrated. This tells the software how to adjust the mount to
correct for any tracking errors. The guide rate (the speed at which the
mount moves to correct errors) and the calibration time (how long the mount is
moved during the test exposures to calibrate the guider) will vary from scope to
scope. I usually use a guide rate of 0.5x with a long-focal-length scope
such as the C14, and 1x with short-focal-length scopes like small refractors.
An appropriate calibration time is then about 5 to 10 seconds. I try to
get the guide star to move around 20-30 pixels. See the Software
Instructions section on using MaxIm DL for
autoguiding for more details on the exact procedure used.
Once the calibration is successful, autoguiding
can begin. A typical guiding exposure is 2-3 seconds long unless the star
is unusually faint (or unusually bright, which is uncommon). For this
setup, only LRGB filters were used, so the guiding exposures were short.
In situations where narrowband filters such as Hydrogen-Alpha are used, guiding
exposures can be 10-20 seconds. I let the autoguider run for a full minute
or so (20-30 guiding corrections) to be sure everything is working fine and to
let the system settle in. You will see autoguider errors listed. The
maximum allowable value depends primarily on the focal length of the scope and
the pixel size of the CCD. An error of 2 pixels is only 1 arcsecond on a
C14 at f/7 with an ST-10XME. The same 1 arcsecond error on a
500mm-focal-length refractor is only 0.36 pixels. Once I see that the
guiding errors are not too large, I begin the exposures.
Luminance Exposure
Once the software is autoguiding the mount, the
exposures can begin. Luminance exposures are binned 1x1; in other words
they are shot at full resolution. Multiple images are acquired and then
stacked together to produce a final image. It is preferable, for example,
to take three 5-minute exposures versus one 15-minute exposure. So two
things need to be determined: the length of each exposure and the total number
of exposures. The Imaging Theory section has a discussion on ideal
exposure times and subframes, full of all sorts of delightful equations.
On the other hand, you could do what I do: pick based on attention span.
For this exposure sequence I took six 10-minute images. 10 minutes is
about how long I can go without checking in on the computer to make sure
everything is working right. Also, if things go bad during an exposure,
I'm only out 10 minutes. It turns out that mathematically this is not too
far off from the ideal exposure time (it's actually a bit longer than ideal).
Six total exposures is a good number for stacking. More is better but
there is a point of diminishing returns. Also, as will be seen below,
this produced a total imaging time of two hours for both luminance and RGB, and
that's about how long I can stand to image one part of the sky. Someone
with more patience could likely have imaged longer and captured even more
detail.
In MaxIm DL a Sequence was setup to capture six
600-second exposures. The sequence was begun and each image was briefly
checked after downloading to make sure everything looked okay.
Color Images
After the luminance sequence was finished, the
autoguider was stopped. This was done to prevent losing the guide star
while the filters were changed. The Red filter was selected and the guide
star reacquired. Calibration does not need to be done after changing
filters unless a longer exposure is required. If the guide star is
reasonably bright during the luminance exposure, there will be no need to
lengthen the exposure for the color filters.
It is best to check focus when changing to the
red filter (and again for each other color). The lenses in the optical
system may cause each color to focus slightly differently. Also, after an
hour of luminance imaging, focus may have changed due to the temperature
dropping.
Selecting RGB Exposure Times
The color images were binned 2x2, or shot at
half resolution. This increases the sensitivity of the camera and allows
shorter exposures. Since the final resolution of the image will come from
the luminance image, shooting low-resolution color frames is fine. The
trick is to get approximately the same amount of data in both the luminance and
RGB images. This is simply a matter of determining equivalent exposure
times, which is easy enough to do.
An exposure through a red, green, or blue filter
will normally be about 1/3 of the same exposure through a clear filter. In
other words, a 10-minute exposure through the clear filter would require a
30-minute exposure through the red filter for equivalent depth. But, since
the camera is binned 2x2 for the color images, the sensitivity has increased
fourfold. This means the equivalent exposure is 30/4 or 7.5 minutes, or
450 seconds. In the interest of my short attention span, I rounded down
and took 400-second exposures for the RGB images.
|
Luminance Exposure |
RGB Equivalent Binned 2x2 |
RGB Equivalent Binned 3x3 |
|
60 seconds |
45 seconds |
20 seconds |
|
90 seconds |
68 seconds |
30 seconds |
|
120 seconds |
90 seconds |
40 seconds |
|
150 seconds |
113 seconds |
50 seconds |
|
300 seconds |
225 seconds |
100 seconds |
|
500 seconds |
375 seconds |
167 seconds |
|
600 seconds |
450 seconds |
200 seconds |
Since most of the detail will be obtained from the luminance exposure, it is
not necessary to take so many individual frames for the RGB images. 3 each
is sufficient to remove most of the noise from the images. Any remaining
noise can be eliminated during post-processing.
The total exposure sequence for the NGC253 image was then:
After refocusing, the autoguider was restarted and the Red sequence was
captured. The autoguider was again stopped afterwards and the procedure
repeated for the Green and Blue exposure sequences.
Dark Frames
The final step was to take dark frames. Flat frames are also a common
calibration image to take, but I have never used them. The main reason is
that they require either an observatory (which I don't have), being setup before
dark (which never happens), or staying up until sunrise (which I don't do too
often since I have to come into work and write this webpage). Also, I've
never looked at one of my images and said, "You know what this needs? A
flat field." As long as you have darks, you are good to go. If you
want to take flats, they can't hurt. And for imaging systems with a fair
amount of vignetting, flats are highly recommended. If, for example, I was
using an STL-11000 on the C14, there would be some noticeable
vignetting in the corners and flats would be helpful. With the smaller
chip of the ST-10XME, flats are less critical.
It is best to take a sequence of dark frames and combine them to create a
master dark frame. This will eliminate any extraneous noise. More is
better, as always. If I had a permanent observatory and could roll the
roof closed and go to bed, I would probably take ten dark frames. Since I
don't have an observatory, I limit myself to three frames in the interest of
sleep. Remember that you need darks for both the luminance exposure and
the color exposures since differ in length and were taken at different
resolutions. For this image, three 600-second darks binned 1x1 and three
400-second darks binned 2x2 were captured using the Sequence mode of MaxIm DL
(allowing me to go eat snack mix and put my greasy fingers all over my friends'
telescopes). This increases the total time spent imaging to 2 hours 50
minutes.
(An image of NGC7331 was captured the same night using the exact same
exposure times, so the dark frames were used for two images. This helps
make 50 minutes of dark frames seem worth the trouble.)
Summary
At the end of the night, a total of 21 images (including dark frames) had
been captured for NGC253. Image all night, sleep all morning, process
images all afternoon--sounds like a fun day to me! Continue to the next
section for details on how the images were processed.

Processing the Image
Return to Imaging NGC253
Main Page
|