|

Accurate polar alignment is critical for CCD imaging. Field rotation, a
trailing of stars near the edges of an image, can result from improper polar
alignment. Even short exposures such as planetary images can benefit from
good polar alignment as there will be left drift from one image to the next
during a sequence, making later registration of the images easier.
There are three main steps to polar alignment: an initial rough alignment, a
more precise estimate using a polar alignment scope (a highly, highly recommended
accessory), and then, for those seeking the highest level of precision, the
declination drift alignment procedure.
Rough Alignment
This is simply a matter of getting the telescope oriented in basically the
right direction.

Above: The two stars in the end of the Big Dipper's handle
always point toward Polaris, the North Star. This chart shows the position
of the Big Dipper and Cassiopeia in the evening in early summer. As the
seasons progress these constellations will rotate from month to month in a
counterclockwise direction around Polaris.
-
Level the tripod. While this is not strictly necessary, it will be
very helpful for the drift alignment procedure.
-
Make sure the latitude scale on the mount is set correctly. There is no need to be extremely accurate here.
Within 1 or 2 degrees is close enough. Your latitude can be found
on most local maps.
Note: Somewhere around the
$3000 mark, equatorial mounts stop coming with latitude scales. The
reason for this is anybody's guess. If you do not have a latitude scale,
estimating is sufficient for now. Many German equatorial mounts have a clear
view through the polar axis shaft where a polar alignment scope attaches (next
step). Sighting Polaris through this shaft is close enough for rough
alignment.
Polar Alignment Scope
There are more accessories available for CCD imaging than there are stars in
the heavens, but trust us on this one: you need a polar alignment scope.
You don't need it like you need oxygen, but your life will be much easier with a
polar scope. If you are doing a drift alignment, the process is easily 3
times faster (say, 20 minutes instead of an hour) when you use a polar scope.
That's 40 extra minutes of prime CCD imaging time (or 40 more minutes of sleep since
you will finish earlier)!
|
Orthogonality
A possible exception to the use of a
polar scope is if you have a computerized telescope mount which has a polar alignment
procedure built in. Usually this mode involves several iterations of
moving from Polaris to another bright star and making adjustments to the mount
to align with true celestial north. This procedure is not 100% foolproof
because it depends on a difficult-to-pronounce factor called "orthogonality".
Orthogonality is not a small, bird-like dinosaur -- this term actually refers to
whether the optical axis of the telescope is perpendicular to the declination axis of the telescope
mount (and thus perfectly parallel to the polar axis). Perfect orthogonality
is rarely achieved, even with very high-end telescopes. Aligning with
a polar scope, and the drift alignment method, is
independent of orthogonality. |
A polar alignment scope works by showing the offset from Polaris to true
celestial north, which is about 3/4 of a degree. By aligning the mount
using a polar scope, the telescope can track sufficiently for short-exposure
images, or very wide-field exposures. For more accuracy, the drift method
is necessary (see next section).

Above: The view through a typical polar alignment scope.
Polaris is centered, which is not where it needs to be.

Above: Polaris really needs to be in the position shown
above, offset slightly from true celestial north.
The accuracy of this alignment can be improved further by rotating the scope
slightly in right ascension to match up the second alignment star seen at the
top of the diagram above.

Above: Both Polaris and the nearby alignment star are in
their proper positions. A very accurate alignment has been achieved.
Declination Drift Alignment
Once you have read the directions below, you may wish to print out the
Drift Alignment Quick Reference to have with you in
the field. The drift
method of polar alignment requires the use of an illuminated crosshair
eyepiece. A simple double-crosshair (shown below) works perfectly,
although a fancier eyepiece such a Celestron's Micro Guide eyepiece will
work fine as well.
Note: These instructions apply
to telescopes which use diagonals such as SCTs, RCs, and refractors.
Newtonian-style telescopes will need to reverse certain directions which are
noted.

Above: The view through a typical illuminated crosshair
eyepiece.
Azimuth Adjustment
-
The drift alignment requires that you let the telescope track on two
different stars in specific locations in the sky. Watching how the stars drift
relative to the reticle in the crosshair eyepiece tells you how far the mount is
offset from true celestial north and in which direction.
-
Pick a star near the meridian, just
north of the celestial equator (due
south, between about 60°-70° above the horizon from the U.S.). Select a
star that is reasonably bright but not too bright (about magnitude
3-4). Be sure that no other similar stars are in the field of view,
as you do not want to get confused as to which star is which.
-
Aim the telescope to this star. Rotate the diagonal until the
eyepiece is oriented so that you are standing on the north side of the
telescope when looking into the eyepiece. This step is not absolutely
necessary but will make the following procedure easier.
-
The crosshairs of the eyepiece must be aligned with the north-south and east-west
directions. Center the star in the eyepiece. Use the mount's
hand-controller to move the star east and west (roughly left and right) in the
eyepiece. You should see that the star's motion is not perfectly
parallel to the horizontal lines in the eyepiece. Rotate the eyepiece
and check the east-west motion again. Repeat until the crosshairs
are properly aligned.
Above: Rotate the eyepiece so that the crosshairs are
parallel to the east-west motion of the star in the telescope.
-
Once the crosshairs are oriented, place the star on one of
the lines east-west (approximately horizontal) lines. In other words,
the star image should be bisected by one of the horizontal lines as shown
below. Do not place the star between the lines, as it will not provide
enough accuracy for the following steps.

Above: Place star on east-west line.
Newtonian telescope users must reverse the
following directions
-
If the star drifts up, use the mount's azimuth
adjustment knobs to move the mount so that the star appears to move
right
in the field of view.
-
If the star drifts down, use the mount's azimuth adjustment knobs
to move the mount so that the star appears to move left in the field
of view.
|
 |
 |
|
Star Drifts Up |
Adjust Mount to Move Star Right |
-
Use the hand-controller to move the star back onto the horizontal line.
-
Let the star drift again. You should notice that it takes longer
for the star to begin drifting off the line. Repeat the azimuth adjustments,
placing the star back on the crosshair again when finished.
-
Continue letting the star drift and making adjustments until the star takes
about 5 minutes to drift off the line. Again, ignore any left-right
motion. Once the star stays bisected by the line (not just
close to the line) for 5 minutes without any drift, your mount is accurately
aligned in azimuth. Now you just need to adjust the mount in altitude.
Altitude Adjustment
-
Pick a second star in the east, about 20° above the horizon, near the
same declination as your first
star (near the celestial equator). In other words, move the telescope
mostly in right ascension to select the
second star. If there are any obstructions on your eastern horizon,
it is possible to achieve an accurate alignment using a star up to about
50° above the horizon.
-
If you do not have an unobstructed view to the east, a star in the west
can be chosen. You must reverse the adjustments below, however, if
you use a star in the west.
-
Rotate the diagonal so that you are now standing on the south side of the
telescope when looking in the eyepiece. Again, this just makes the
adjustments easier.
-
Orient the crosshairs again as you did above, so that the horizontal crosshairs
are parallel to east-west motion and the vertical crosshairs are parallel
to north-south motion.
-
Place the star on one of the horizontal lines.
-
Let the star drift. You should notice some drift after only a minute
or so unless you initial rough alignment happened to be very good.
The following steps are the same
regardless of the type of telescope used
-
If the star drifts up, use the mount's altitude
adjustment knobs to move the mount so that the star appears to move
down
in the field of view.
-
If the star drifts down, use the mount's altitude adjustment knobs
to move the mount so that the star appears to move up in the field
of view.
|
 |
 |
|
Star Drifts Up |
Adjust Mount to Move Star Down |
-
Use the hand-controller to move the star back onto the horizontal line.
-
Let the star drift again. You should notice that it takes longer
for the star to begin drifting off the line. Repeat the altitude
adjustments, placing the star back on the crosshair again when finished.
-
Continue letting the star drift and making adjustments until the star takes
about 5 minutes to drift off the line. Again, ignore any left-right
motion. Once the star stays bisected by the line (not just
close to the line) for 5 minutes without any drift, your mount is accurately
polar aligned. You are ready to begin imaging the heavens!
Next, Attaching a CCD Camera and Accessories....

Next Page
|